Download Literature |
Get Info or a Quote
Model 221: Mass Flow Instruments
|
Porter Mass Flow products reflect over four decades of experience in the design and manufacture of precision instruments for the measurement and control of gas flow. They incorporate design principles that are simple and straightforward, yet flexible enough to operate under a wide variety of process parameters. The result is flowmeters, flow controllers and control valves that are accurate, reliable and cost-effective solutions for many gas flow applications in the analytical, process, chemical/petrochemical, environmental, biopharmaceutical and research markets.
|
The 200 series is the lastest evolution of the original Porter Analog MFC. With thousands installed worldwide, they are the proven solution when cost effective high performance gas flow control is the goal. The 100 Series Mass Flow Meters are available for applications where flow measurement only is required.
|
|
SPECIFICATIONS Flow Capacity Please consult factory Response Time (per SEMI E17-91 Settling Time) 1 to 2 Seconds Accuracy and Linearity ±1% full scale Repeatability Within ±0.2% full scale at any constant temperature within operating temperature range Rangeability (Control Range) 50; 1 (2%-100% full scale) (accuracy and control) Ambient and Operating Temperature Range -10 to 70 °C ( ±14 to 158 °F) Maximum Operating Pressure 3000 PSIG Temperature Coefficent (per SEMI E18-91 Zero Effect and Span Effect) ±0.05% full scale / °C of zero ±0.05% of reading/ °C of span Mounting Orientation Attitude insensitive
|
Warm-up Time 10 minutes External Electrical Connector Nine (9)- pin Dconnector Weight (approximate) 4.1 lbs Power Supply Requirements (Current consumption <250 mAdc): Voltage output models: +12 (±5%) (0-5 Vdc & 1-5 Vdc flow signal outputs only) or +15 (±10%) Vdc Current loop models: +15 (±5%) or +24 (±15%) Vdc Setpoint Input/Flow Signal Output 0-5 Vdc / 0-5 Vdc (2K ohm minimum load resistance) 0-10 Vdc / 0-10 Vdc (3K ohm minimum load resistance) 1-5 Vdc / 1-5 Vdc (2K ohm minimum load resistance) 4-20 mAdc / 4-20mAdc (refer to load resistance values below) 1-5 Vdc / 4-20 mAdc (refer to load resistance values below) Load resistance values for 4-20 mAsc flow signal output 0-450 ohm for 6.5-15 Vdc loop supply voltage 200-750 ohm for 15-30 Vdc loop supply voltage
|
|
MATERIALS OF CONSTRUCTION Body 316 Stainless Steel Sensor Assembly 316L Stainless Steel Orifice 316 Stainless Steel Valve Components (Wetted) 302 Stainless Steel, 316 Stainless Steel, and Sandvik® 1802 Elastomers (O-rings and Valve Seat) Buna N or Viton® Process Connections 316 Stainless Steel
Sandvik® - AB Sandvik Materials Technology Kalrez®, Viton® - DuPont Dow Elastomers L.L.C. Specifications subject to change
|
ORDERING INFORMATION To order, please specify • Model number • Type of output signal • Elastomer material • Process connection size and type • Flow capacity • Gas type • Operating temperature • Inlet (supply) pressure • Outlet pressure • Calibration standard (i.e. 0°C, 20°C, 21.1°C or 25°C) • Additional accessories required
|
Download the complete PDF for more details and ordering information
Request info or a quote on this product