| PORTER Liquid Mass Flow - Specifications |
D2000i Mass Flow
Click
Here to download specifications and ordering form.
Specifications: Response Time: 1 seconds to within 5% of setpoint Accuracy and Linearity: ± 1% of reading Repeatability: <±0.2% of rate Rangeability (Control Range): 20:1 (5%-100% full scale) (accuracy & control) Maximum Operating Temperature: 40°C (104°F) Maximum Operating Pressure: Temperature Coefficient: < 0.1%/°C Setpoint Input/Flow Signal Output: 0-5 Vdc ( 2k ohm minimum load resistance for flow signal output) Power Supply Requirements: (Current Consumption < 200 mAdc): +15 (±10%) Vdc Leak Integrity: 1 x 10-9 atm. cc/sec.He Warm-up Time: 20 minutes Process Connections: 1/8" Male VCR or 1/8" Swagelok Materials of Construction- (wetted
surfaces): Control Valve: Normally open piezoelectric-actuated External Electrical Connector: Nine (9)-pin D-connector Internal Volume: < 1 ml Weight: Less than 1.5 lbs
|
![]() Flow Capacity: Following are the available flow ranges based on isopropyl alcohol (IPA). To determine the flow range for process fluids other than IPA, multiply the IPA flow range by the appropriate correction factor listed below. IPA Flow Ranges (in ml/min):
.003-0.14 Correction Factors for Process Fluids Other than IPA: Liquid Correction Factor |
Contact us to size fluids
not listed or when operating parameters are questionable |